List of program instructions – Metrohm IC Net 2.3 User Manual
Page 90

6 Devices
IC Net 2.3
80
<Add>
Add new program instruction.
<Delete>
Delete selected program instruction.
<Verify>
Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
subpage:
Instruction Parameter entry
Meaning
Remote
0, 1, *
Set remote output lines 1...8 to
the desired values. For entry of
the first value, enter
1
,
0
or
*
. For
entry of the other values, move
the cursor in front of the value to
be changed and enter
1
,
0
or
*
.
Range
100, 200, 5000
µ
S/cm
1, 2, 5, 10 mS/cm
Set measuring range to the se-
lected value.
FS 0.05
µ
S/cm ...
10 mS/cm
Set full scale range to the se-
lected value (depends on
Range
setting).
Zero on,
off
Switch on or off the auto-zero
function.
Polarity +,
–
Reset the polarity of the output
signal.
Mark
Trigger marking signal.
ValveA fill,
inject
Switch injection valve A to "
inject
"
or "
fill
" position.
ValveB fill,
inject
Switch injection valve B to "
inject
"
or "
fill
" position.
Suppressor
Switch suppressor module to the
next position.
StartDA
Start data acquisition.