Controls – Metrohm IC Net 2.3 User Manual
Page 260

6 Devices
IC Net 2.3
250
Controls
On the subtab
Controls
injection valve, peristaltic pump, high-
pressure pump, column heating and remote lines can be manu-
ally controlled. In addition, the parameters for the high-pressure
pump and column heating temperature can also be set:
Valve
Injection
valve
<Inject>
Switch injection valve to "Inject" position.
<Fill>
Switch injection valve to "Fill" position.
Peristaltic pump
Peristaltic
pump
<On>
Start peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).
<Off>
Stop peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).
IC pump
High pressure pump
<On>
Start high pressure pump.
<Off>
Stop high pressure pump.
Flow, mL/min
Set value for flow rate of IC pump.
Entry range:
0.20
...
2.50 mL/min
Pressure max, MPa
Value for maximum switch-off pressure for
high-pressure pump. Once it has been set
this value will be monitored, even if there is
no connection to the PC (units depend on
the settings made under
IC NET / Global
preferences
).
Entry range:
0.0
...
35.0 MPa
If the
Flow
is higher than 1.5 mL/min, the
defined
Pressure max
must not exceed