Metrohm 850 Professional IC Cation - LP-Gradient User Manual

Page 23

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

850 Professional IC Cation – LP Gradient

■■■■■■■■

13

■

Connect the leak sensor (see "Connecting the leak sensor ", page
22)
.

■

Connect drainage tubings (see "Installing drainage tubings",
page 25)
.

3 Installing the eluent path

■

Equip the eluent aspiration tubing and connect with the eluent
bottle (see Chapter 3.8.1, page 28).

■

Instead of the column, connect the 6.2744.040 coupling with a
6.2744.070 short pressure screw at the end of the pre-installed
column inlet capillary.

■

Connect the detector inlet capillary with a 6.2744.070 short pres-
sure screw to the 6.2744.040 coupling.

4 Installing the sample path

Note

The sample degasser does not have to be connected. We recom-
mend the usage of the sample degasser only if the sample matrix
requires it.

■

Guide the 6.1803.040 sample aspiration capillary connected to
the sample input of the injection valve out of the instrument
through a capillary feed-through and connect it with the Sample
Processor or

■

If the sample degasser is used: Connect the sample aspiration
capillary with a 6.2744.090 long pressure screw to the output of
the sample degasser (see Chapter 3.14, page 44).
Connect a part of the 6.1803.040 sample aspirating capillary with
a 6.2744.090 long pressure screw to the input of the sample
degasser. Guide the other end out of the instrument through a
capillary feed-through and connect it to the Sample Processor (see
Sample Processor manual)
.

5 Expanding the instrument to low pressure gradient

See Chapter 3.10, Page 34

■

Loosen the 6.1834.090 connection tubing (1-5) from the output
of the eluent degasser.

■

Dismount the coupling (20-9) and pump head inlet capillary
(20-7) of the high pressure pump.