3.19 Separation column

■■■■■■■■■■■■■■■■■■■■■■

54

■■■■■■■■

850 Professional IC Cation – LP Gradient

each column. You can request detailed information on special IC applica-
tions in the corresponding "Application Bulletins" or "Application
Notes
", available in the Internet at

http://www.metrohm.com

in the

Applications area or via the Metrohm agent responsible free of charge.

Note

The separation column may only be installed after the initial start-up
(see Chapter 4.1, page 55) of the instrument. Until then, use the
6.2744.040 coupling instead of the guard and separation column.

Connecting and rinsing the separation column

1 Connecting the separation column

Caution

When using the columns, always ensure that these are correctly
inserted corresponding to the flow direction indicated.

■

Remove stoppers from the separation column.

■

Connect the lower end of the separation column with a
6.2744.070 short PEEK pressure screw to the outlet capillary of
the guard column (if used), or to the column input capillary .

2 Rinsing the separation column

■

Place beaker under the outlet end of the separation column.

■

Set the flow rate of the high pressure pump to the value suitable
for the separation column selected.

■

Start the high pressure pump and rinse the separation column
approx. 10 minutes with eluent.

■

Switch off the high pressure pump again.

3 Mounting the separation column

■

Hang separation column with chip into the column holder.

Note

The iColumns are equipped with a chip on which their operating data is
saved. The chip has to be hooked into the chip holder provided for this
so that the column recognition can function.