7.3 Method setup

IC Net 2.0

107

7.3.3

Filters

Tab

Filters

of

Method setup

window with parameters for noise fil-

tration of raw data. Three different filtration algorithms are avail-

able to reduce noise and increase apparent signal-to-noise ratio.

Spikes

Spikes filter. The Spikes filter smoothes

the first and last points of the chroma-

togram and the points identified as spikes.

The spike is exchanged with half of the sum

of two neighboring signal values.

Slit Median

Median filter. Performs filtration if non-zero

smoothing degree is entered. Smoothing

degree should be a natural number. The

values within the window are sorted by in-

creasing response level and the response

corresponding to the middle of the window

is replaced with the value in the middle of a

sorted array. This method affects chroma-

tographic peaks in minimal extent, im-

proves baseline and eliminates single-point

spikes. In this case the spike will be re-

placed with one of the neighboring points.

Slit Gaussian

Gauss filter. Performs filtration if non-zero

smoothing degree is entered. The sum of

all window points with Gaussian weights

distribution is calculated and is used as a

new raw data value. The peaks become a

bit smaller and wider after smoothing, but

their area does not change.

Raw data points

Number of data points per chromatogram

(read-only).

Smallest peak has ...

Number of data points per smallest peak

on the chromatogram (read-only). This in-

formation is valuable to perform the

Com-

press

procedure (data compression).