Metrohm IC Net 2.0 User Manual

Page 180

Advertising
background image

10 Batch reprocessing

IC Net 2.0

172

bration parameters (component table, con-

centration table and calibration curve) are

applied to all sample runs.

Apply final calibration to all reprocessed files

Apply the updated calibration to all calibra-

tion and sample runs. If this option is

switched off, the calibration stored in the

first chromatogram is used for all other

chromatograms.

Forget calibration points before reprocessing

Forget old calibration points of the calibra-

tion curve and perform a new calibration

using all calibration runs of the batch re-

processing queue. In this case each cali-

bration chromatogram adds a new point to

the calibration curve. If this option is

switched off, the calibration curve stored in

the first chromatogram remains active.

Each further calibration chromatogram in

the batch reprocessing queue adds a new

point to this calibration curve.

Recalculate only

Enable recalculation of chromatograms

with values for

Volume

,

Dilution

,

Amount

and

Internal standard amount

entered in the batch

reprocessing table.

The recalculation is done automatically if
the

Reintegrate

and/or

Recalculate

options

are enabled. If the

Recalculate only

option

is enabled, the

Reintegrate

and

Recalculate

options are disabled automatically.

Change passport

If this option is enabled, those parameters

of the passport which have been changed

after clicking the

<Edit passport>

button are

applied to all chromatograms.

<Edit passport>

Open the

Passport

window for modification

of the passport parameters.

Only some of the passport parameters
can be modified. The passport parame-
ters

Ident

,

Sample Info 1

and

Sample Info 2

entered in the batch reprocessing table
are overwritten if these values are modi-
fied in the

Passport

window.