Metrohm IC Net 2.0 User Manual

Page 29

Advertising
background image

4.3 System functions

IC Net 2.0

21

SYSTEM / Control / Disconnect from workplace

Disconnect selected system from the workplace. If a system is

disconnected, manual control is not available for this system, but

all other system settings can be modified and saved.

Start deter-

mination

and

Startup hardware

is only possible if no other system

from the same system folder is connected. In this case, the sys-

tem is automatically connected.

4.3.2

Start/stop hardware and record baseline

SYSTEM / Control / Startup hardware (Measure Baseline)

With this menu item the following actions are executed automati-

cally:

Send

System startup values

to the devices.

Start Metrohm solvent delivery systems and Bischoff solvent

delivery systems.

Start Metrohm additional modules and Bischoff additional

modules.

Start Bischoff detectors.

Switch Metrohm Autosamplers and Bischoff Autosamplers to

remote control.

Start recording of the measurement signal using the method of

the connected system. Independently of the set chromatogram

Duration

, the measurement signal is recorded until the data

acquisition is stopped if a new determination is started with

Start determination

or by clicking the icon of the chroma-

togram window. In this case the user is asked if the recorded

baseline should be saved or not.

SYSTEM / Control / Shutdown hardware

With this menu item the devices are stopped as follows:

Stop Metrohm solvent delivery systems and Bischoff solvent

delivery systems.

Stop Metrohm additional modules and Bischoff additional

modulesVariotherm.

Stop Bischoff detectors.

Switch Metrohm Autosamplers and Bischoff Autosamplers to

local control.

Stop running determination.

Stop active sample queue.

SYSTEM / Control / Stop data acquisition

Stop recording of the baseline.