5.4 761 Compact IC
IC Net 2.0
43
IC pump
High-pressure pump.
<On>
Start high-pressure pump.
<Off>
Stop high-pressure pump.
Peristaltic pump
Peristaltic pump.
<On>
Start peristaltic pump.
<Off>
Stop peristaltic pump.
System startup values
The system startup values are sent and
applied to the 761 Compact IC each time
the system is connected, a determination is
started, or the values are sent manually
with
<Send to unit>
.
Flow, mL/min
Startup value for flow rate of the high-
pressure pump.
Entry range:
0.20
...
2.50 mL/min
Pressure max, MPa
Startup value for maximum pressure limit
for high-pressure pump. This limit is con-
trolled even without connection to the PC.
Entry range:
0.0
...
25.0 MPa
Pressure min, MPa
Startup value for minimum pressure limit for
high-pressure pump. This limit is controlled
even without connection to the PC.
Entry range:
0.0
...
25.0
MPa
Full scale,
µµS/cm
Startup value for full scale range.
Selection:
50
,
250
,
1000
µµS/cm
Remote lines
Startup value for remote line settings 1
…
8.
Selection:
0
,
1
Initial
The
Initial
tab of the
761 Compact IC
window for settings of system
startup values is only available for a disconnected system.
Flow, mL/min
Startup value for flow rate of the high-
pressure pump.
Entry range:
0.20
...
2.50 mL/min