5.4 761 Compact IC
IC Net 2.0
45
Second column
Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the
Remote configura-
tion
tab.
Third column
Parameter for program instruction (see
below).
ENABLED
Enable program start (a disabled program
is not started).
<Add>
Add new program instruction.
<Delete>
Delete selected program instruction.
<Verify>
Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
subpage:
Instruction
Parameter entry
Meaning
Valve
Inject, Fill
Switch injection valve to "INJECT"
or "FILL" position.
FullScale
50, 250, 1000
µµS/cm
Set full scale range to the se-
lected value.
ICPump
on, off
Switch on or off the high-pressure
pump.
Flow
0.2 ... 2.5 mL/min
Set flow rate of the high-pressure
pump to the desired value.
Pmax
0.0 ... 25.0 MPa
Set maximum pressure limit for
the high-pressure pump to the de-
sired value.
Pmin
0.0 ... 25.0 MPa
Set minimum pressure limit for
the high-pressure pump to the de-
sired value.
Remote
0, 1, *, p
Set remote output lines 1...8 to
the desired values. For entry of the
first value, enter
1
,
0
,
p
or
*
. For en-
try of the other values, move the
cursor in front of the value to be
changed and enter
1
,
0
,
p
or
*
.