3 precolumn glass cartridges with cartridge holder – Metrohm 790 Personal IC User Manual

Page 41

Advertising
background image

2 Installation

790 Personal IC

32

2.7.3

Precolumn glass cartridges with cartridge holder

The precolumn glass cartridge METROSEP Anion Dual 1 (6.1006.030)

is inserted in the 6.2828.010 Precolumn cartridge holder as follows (see

Fig. 12):

1 Insert cartridge

• Remove end fittings 64

64

64

64 from screw caps 65

65

65

65.

• Insert precolumn cartridge 67

67

67

67 in sleeve 66

66

66

66 (mark flow direc-

tion of the column cartridge on sleeve, the column cartridge

should always be operated in the same flow direction).

• Screw both screw caps 65

65

65

65 loosely to sleeve 66

66

66

66 by hand.

• Screw both end fittings 64

64

64

64 into screw caps 65

65

65

65 so that their

capillary ends are seated in the PTFE seals of the cartridge.

• Tighten both screw caps 65

65

65

65 by hand.

• Tighten both end fittings 64

64

64

64 firmly using a screwdriver.

2 Connect precolumn

• Provide the column connection capillary 20

20

20

20 mounted on the

injection valve with a PEEK compression fitting 46

46

46

46 (see sec-

tion 2.3.4) and screw it tightly onto end fitting 64

64

64

64 on the inlet

side of the precolumn.

• Cut off as short a piece as possible from PEEK capillary 48

48

48

48

and fit this with two PEEK compression fittings 46

46

46

46 (see sec-

tion 2.3.4).

• Screw prepared capillary 48

48

48

48 to end fitting 64

64

64

64 on the other end

of the precolumn.

3 Rinse precolumn

• Place a beaker beneath the outlet capillary of the precolumn.
• Open software window for manual system control.
• If necessary, modify

Flow rate

to the value suited for the

inserted separating column and click on

<Send to unit>

to

send this value to the 790 Personal IC.

• Switch on high-pressure pump (

IC pump

) by clicking

<On>

and

rinse precolumn with eluent for ca. 10 min.

• Switch off high-pressure pump by clicking

<Off>

.


The rinsed precolumn can be connected directly to the separating

column instead of by means of a PEEK capillary. Proceed as follows:
• Screw off PEEK capillary from end fitting 64

64

64

64 of the precolumn.

• Screw off end fitting 64

64

64

64 from screw cap 65

65

65

65 of the precolumn.

• Screw connection piece 68

68

68

68 onto screw cap 65

65

65

65.

• Screw screw cap 69

69

69

69 of separating column 63

63

63

63 onto connection

piece 68

68

68

68.