Metrohm 790 Personal IC User Manual

Page 57

Advertising
background image

2 Installation

790 Personal IC

48

• Cut PEEK capillary 18

18

18

18 installed at the 766 IC Sample Proces-

sor (see 766 Instructions for Use) to the desired length.

• Pull the free end of PEEK capillary tubing 18

18

18

18 through the

feedthrough 4

4

4

4 of the 790 Personal IC and screw it onto con-

nection "1" of injection valve 22

22

22

22 with the help of a 6.2744.010

PEEK compression fitting.

3 Tubing connection injection valve – waste

• Remove coupling 24

24

24

24 and 6.2744.010 Compression fitting

from syringe tubing 25

25

25

25.

• Lead free end of syringe tubing 25

25

25

25 to the waste container and

fix in place.

4 Settings at the 766 IC Sample Processor

• For operation with the 790 Personal IC it is recommended

that the following program "790" is used with the 766 IC

Sample Processor:

parameters
method 790
number of samples: Rack

>start sequence
1 CTL:Rm: INIT
>sample sequence
1 SCN:Rm : Wait1

2 MOVE 1 : sample
3 LIFT: 1 : work mm

4 SCN:Rm : Wait1

5 PUMP 1.1 : 120 s

>final sequence
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------

− Method name

− Number of samples to be processed

(entire sample rack)


− Initialize remote interface


− Waiting until 790 Personal IC sends

signal on remote line 3

− Move needle to sample position

− Place lift with needle to working

position

− Waiting until 790 Personal IC sends

signal on remote line 3

− Fill sample loop with sample during

120 s


− Settings for changer functions




− Reaction to manual stop

5 Settings in the «790 Personal IC» program

• A time program must be drawn up for the selected system

which first produces an impulse at remote lead 3 to start

sample changing on the 766 IC Sample Processor and then

switches the injection valve to the "Fill" position. A further im-

pulse is then produced to start filling the sample loop within

120 s. At the end, the injection valve is switched to the "Inject"

position, which also starts data acquisition.