Metrohm 790 Personal IC User Manual

Page 90

Advertising
background image

4.3 Systems

790 Personal IC

81

Time (1

st

column)

Time at which program instruction is applied.

Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruction is
applied together with the last instruction with
time entry.

Command (2

nd

column)

Program instruction (see List of program in-

structions).

Parameter (3

rd

column)

Parameter for program instruction (see List of

program instructions).

ENABLED

Enable program start (a disabled program is

not started).

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are dis-

played if program is faulty).

List of program instructions
The following program instructions can be added to the time program

on the

Program

subpage:

Instruction Parameter

entry

Meaning

Valve Inject,

Fill

Switch

injection valve to "INJECT" or

"FILL" position.

ICPump on,

off

Switch

on or off the high-pressure

pump.

Flow

0.2 ... 2.5 mL/min

Set

flow rate of the high-pressure

pump to the desired value.

Pmax

0.0 ... 25.0 MPa

Set

maximum pressure limit for the

high-pressure pump to the desired
value.

Pmin

0.0 ... 25.0 MPa

Set

minimum pressure limit for the

high-pressure pump to the desired
value.

Remote

Pulse output (pulse length = 200 ms)
on remote output line 3 for sample
changer control.

Program

END, RESET

The

END

flag can be used to end a

program, especially if the program time
should be longer than the chroma-
togram duration. Additional steps after
this flag are not allowed. The

RESET

flag

is used to reset the parameters to the
system startup values.