4.2 Remote control commands

756/831 KF Coulometer, Instructions of Use

71

&Setup

Object Description

Input

range

Reference

& Root

:

├

Setup

Settings for the operating mode

├

.Comport

Output of automatic info

1,2,1&2

4.2.2.85.

├

.Keycode

Send key code

ON, OFF

4.2.2.86.

├

.Tree

Sending format of path info

│ ├

.Short

Short format of path

ON, OFF

4.2.2.87.

│ ├

.ChangedOnly

Paths of modified nodes only

ON, OFF ditto

│

├

.Trace

Message on changed values

ON, OFF

4.2.2.88.

│

├

.Lock

Lock key functions

│ ├

.Keyboard

Lock all keyboard keys

ON, OFF

4.2.2.89.

│ ├

.Config

Lock <CONFIG> key

ON, OFF

ditto

│ ├

.Parameter

Lock <PARAM> key

ON, OFF

ditto

│ ├

.SmplData

Lock <SMPL DATA> key

ON, OFF

ditto

│ ├

.UserMeth Lock

functions

│ │ ├

.Recall Lock

“loading”

ON,

OFF

ditto

│ │ ├

.Store Lock

“saving”

ON,

OFF

ditto

│ │ ├

.Delete Lock

“deletion”

ON,

OFF

ditto

│ ├

.Exchange

Lock <EXCH> key

ON, OFF

ditto

│ └

.Display

Lock display function

ON, OFF

ditto

│

├

.Mode

Setting waiting intervals

│

├

.StartWait

Waiting time after start

ON, OFF

4.2.2.90.

│

├

.FinWait

Waiting time after run

ON, OFF

ditto

│

├

.SendMeas

Automatic sending of measured values

│

├

.SendStatus Connect/disconnect

sending ON,

OFF

4.2.2.91.

│

├

.Interval Time

interval

0.4...4...16200,

ditto

│

MPList

│

├

.Select Selection

Assembly,

Titrator

4.2.2.92.

│

├

.Assembly From

assembly

│ │ ├

.CyclNo Cycle

number

ON,

OFF 4.2.2.93.

│ │ ├

.I

Total charge (mA·s)

ON, OFF ditto

│ │ ├

.Meas

Measured indicator voltage

ON, OFF ditto

│ │ ├

.Pot

Voltage at generator electrode

ON, OFF ditto

│ │ ├

.IPulse

I of current pulse

ON, OFF ditto

│ │ ├

.Bur Connected

buret

│

│ │ ├

.V

Volume of dosing unit

ON, OFF ditto

│ ├

.Titrator From

Titrator

│ │ ├

.CyclNo Cycle

number

ON,

OFF 4.2.2.94.

│ │ ├

.Water

Mass of water

ON, OFF ditto

│ │ ├

.Meas

Measured indicator voltage

ON, OFF ditto

│ │ ├

.dWaterdt

Drift or rate

ON, OFF ditto

│ │ ├

.I

Total charge (mA·s)

ON, OFF ditto

│ │ ├

.Pot

Voltage at generator electrode

ON, OFF ditto

│ │ ├

.IPulse

I of current pulse

ON, OFF ditto

│