4.2 Remote control commands

756/831 KF Coulometer, Instructions of Use

83

.Method: Method used to process the sample, from the method memory or from the

card.

.Id:

The identifications Id1...Id3 can also be used as sample-specific calcula-

tion constants C21...C23 in formulas.

.UnitSmpl: If "no unit" is desired for the sample size, the blank string must be en-

tered.

.C24, .C25: Results which have been assigned to C24 and C25.

.Mark:

Mark of the silo line: "*"=deleted line, "+"=line which is worked off, "-

"= line which is worked off and not valid for silo calculations (deleted),

"/" last worked-off line, where recalculation can still be done. Silo lines

which have been worked off are "read only".

4.2.2.58. SmplData.ONSilo.DelLine $G

SmplData.ONSilo.DelLine.LineNum 1...255, OFF

Deletion of a silo line. The line # is deleted with &SmplData.ONSilo.DelLine $G. If a

formerly deleted line is edited again, it becomes valid (function "undelete").

4.2.2.59. SmplData.ONSilo.DelAll $G

Deletes the entire silo memory. Must be triggered with $G.

4.2.2.60. SmplData.ONSilo.CycleLines ON, OFF

Silo data cycling.

With "ON", executed lines are copied to the next free silo lines, see page 49. Exercise

caution if you edit the silo memory during the determinations!

4.2.2.61. SmplData.ONSilo.SaveLines ON, OFF

Silo lines are not deleted when they are worked off. Assigned results are stored as

C24 and C25. "Save lines" can only be set to "ON" if the silo is completely empty.

Delete the silo, see 4.2.2.60.

4.2.2.62. HotKey.User.Name

up to 10 ASCII characters

HotKey.User.Delete $G

HotKey.User.Delete.Name

up to 10 ASCII characters

HotKey.User.DelAll $G

HotKey.User.List.1.Name read

only

Management of user names.

.Name:

Input of user names.

.Delete,Name: Deletes selected user name with &HotKey.User.Delete $G.

.List:

List of all user names.

4.2.2.63. Info.Report $G

Info.Report.Select

result, water crv, rate crv, meas crv, comb,

mplist, param, calc, C-fmla, def, statistics, smpl data,

silo, scalc full, scalc srt, config, user method, all, ff

$G sends the selected report to the COM which is set in

&Config.PeriphUnit.RepToComport: