5 setting up tubings and tips, Setting up tubings and tips – Metrohm 870 KF Titrino plus Installation User Manual
Page 19

■■■■■■■■■■■■■■■■■■■■■■
2 Installation
870 KF Titrino plus
■■■■■■■■
13
2.5
Setting up tubings and tips
Setting up the aspiration and dosing tip
1
2
3
4
5
6.2730.030
6.1543.110
6.1543.120
1 Place the three stoppers 6.2730.030 (including O-rings, but without
nipples) in the rear openings of the vessel lid.
2 Screw on the dosing tip 6.1543.110 to the M8 PTFE tubing of the
solvent bottle (amber glass).
3 Insert the dosing tip through the stopper in the middle rear opening
of the vessel lid.
4 Screw the aspiration tip 6.1543.120 onto the M8 PTFE tubing of the
aspiration bottle (clear glass).
5 Insert the aspiration tip through the stopper in the right rear opening
of the vessel lid.
The end of the tip must touch the bottom of the vessel, but it must
not inhibit the action of the stirring bar.
As an alternative to the aspiration tip 6.1543.120, the aspiration tip
6.1821.070 can also be used with an M10 screw nipple.