3.2 Carrying out titrations
■■■■■■■■■■■■■■■■■■■■■■
24
■■■■■■■■
870 KF Titrino plus
■
Select the parameter Solution and confirm with [OK].
The list of the configured solutions is displayed:
■
Select the newly configured solution and apply with [OK].
3 Adjust the calculation
■
With [BACK], switch to the parameters dialog.
■
Select the menu item Calculation and open the corresponding
dialog with [OK].
The calculation formula is defined for every method template and
cannot be modified.
■
Select the parameter Factor (FCT) and open the editing dialog
with [OK].
■
Enter the water content of the water standard used in mg/g (see
certificate of the water standard).
■
Apply the value with Accept or [BACK].
Note
In the manual you will find details as to how the conversion factor
FCT is to be adjusted if you do not use a water standard to carry
out the titer determination.
4 Activate the statistics
■
With [BACK], switch to the parameters dialog.