9 installing the pulsation absorber, Installing the pulsation absorber, Figure 10 – Metrohm 930 Compact IC Flex Oven/SeS/PP User Manual

Page 33: Inline filter, Figure 11, Pulsation absorber

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

930 Compact IC Flex Oven/SeS/PP (2.930.2500)

■■■■■■■■

25

Figure 10

Inline filter

1

Input capillary
Connected to the purge valve.

2

Output capillary
Connected to the pulsation absorber.

The inline filter is completely connected. No installation work is required.

3.9

Installing the pulsation absorber

The pulsation absorber is installed between the high-pressure pump and
the injection valve. It protects the separation column from damage caused
by pressure fluctuations such as those when switching the injection valve,
and reduces interfering pulsations during highly sensitive measurements.

Figure 11

Pulsation absorber

1

Connection capillary
Connection to inline filter.

2

PEEK pressure screws, short
(6.2744.070)

3

Pulsation absorber (6.2620.150)

4

Holder for pulsation absorber

5

Connection capillary
Connection to injection valve.

Advertising