background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

930 Compact IC Flex Oven/SeS/PP (2.930.2500)

■■■■■■■■

49

4 Deaerating the high-pressure pump

■

Push the end of the purging needle (6.2816.040) over the end of
the deaerating capillary on the purge valve.

■

Insert the syringe (6.2816.020) in the Luer connector of the purg-
ing needle.

■

Open the purge valve using the rotary knob (approx. ½ turn).

■

Switch on the high-pressure pump in MagIC Net.

■

Use the syringe to aspirate eluent until there are no more air bub-
bles in the aspirated eluent.

■

Switch off the high-pressure pump in MagIC Net.

■

Seal the purge valve using the rotary knob.

■

Remove the syringe from the purging needle.

■

Pull the purging needle out of the deaerating capillary.

5 Setting the contact pressure of the peristaltic pump

■

Set the contact pressure of the peristaltic pump (see "Setting the
contact pressure correctly", page 40)
.

6 Rinsing the instrument without columns

■

Rinse the instrument (without columns) with eluent for 5 minutes.