Metrohm 930 Compact IC Flex Oven/SeS/PP User Manual

Page 47

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

930 Compact IC Flex Oven/SeS/PP (2.930.2500)

■■■■■■■■

39

■

Install the pump tubing connection with safety device and filter
(6.2744.180) at the pump tubing outlet:

– Push the locknut onto the pump tubing.
– Push the appropriate adapter onto the pump tubing.
– Place the tubing olive with the filter holder into the pump

tubing so that the pump tubing is firmly in place; push the
end of the pump tubing over at least the second notch of
the olive.

– Tighten it using the union nut.

2 Removing the tubing cartridge

■

Press in the tubing cartridge's snap-action lever.

■

Tilt the tubing cartridge upwards.

■

Unhook the tubing cartridge from the mounting bolt.

3 Inserting the pump tubing

■

Press the tubing cartridge's contact pressure lever down all the
way.

■

Place the pump tubing in the tubing cartridge. Fit the tubing car-
tridge between two stoppers. The stoppers must snap into the
corresponding holder of the tubing cartridge.

4 Inserting the tubing cartridge

■

Hang the tubing cartridge in the mounting bolt and press it in the
cartridge holder until you hear the snap-action lever snap in.

Setting the flow rate
The flow rate of the peristaltic pump depends on many factors:

■

The inner diameter of the pump tubing

■

The rotational speed of the drive

■

The contact pressure of the tubing cartridge

NOTE

Pieces of pump tubing are consumables. The lifetime of the pump tub-
ing depends on the contact pressure, among other factors.

Advertising