2.16 Silo memory for sample data

799 GPT Titrino

97

Key <SMPL DATA> with the silo memory switched on

Sample data can be entered into the silo memory with

key <SMPL DATA>.

edit silo lines:

Entering sample data into the silo memory.

delete silo lines:

Deletes single silo lines.

delete all silo lines:

Deletes the whole silo memory.

The display texts of the Titrino are shown to the left.

The values are the default values.

>edit silo lines

silo line 1




method:





id#1 or C21

id#2 or C22

id#3 or C23


smpl size 1.0 g



smpl unit: g

Input for silo memory

Silo line (1...255)

The next free line is displayed automatically. Lines

already occupied can be corrected.

Method with which the sample is processed (method

name from the method memory)

If no method name has been entered, the sample is

processed with the method in the working memory.

Selection of the method with <←/→>.

Sample identification 1..3 or sample specific calculation

variables C21...C23 (up to 8 characters)

Sample size (6-digit number: ±X.XXXXX)

The method specific limits are tested on result

calculation.

Unit of sample size (g, mg, mL, µL, pc, no unit or up to

5 characters)

Select unit with <←/→>.

>delete silo lines

delete line n OFF

Delete individual silo lines

Line number of the line to be deleted(1...255, OFF)

<CLEAR> sets "OFF".

Deleted lines remain in the silo memory. Access is

blocked during the processing. To show that a line has

been deleted, they are marked with "*". The symbol *

indicates that the line has been deleted.

Deleted lines can be reactivated if the appropriate line

is re-edited.

SMPL

DATA

smpl data

>edit silo lines

>delete silo lines

>delete all silo lines

cycle lines: OFF

save lines: OFF