4.1. Error and special messages
799 GPT Titrino
192
Error messages in connection with the data transfer
If neither a computer nor a printer is attached, the report output at the end of the titration
must be switched off:
Receive errors:
error 36
Parity
Exit: <QUIT> and set corresponding quantity the
same on both instruments
error 37
Stop
bit
Exit: <QUIT> and set corresponding quantity the
same on both instruments
error 38
Overrun error. At least 1 character could not be read.
Exit: <QUIT>
error 39
Overflow of the receive buffer of the Titrino (> 128
characters).
Exit: <QUIT>
Send errors:
error 42
CTS=OFF
Handshake unsatisfactory for more than 1 s.
Exit: <QUIT> Is the receiver switched on and ready
to receive?
error 43
The transmission of the Titrino has been interrupted
with XOFF for at least 6 s.
Exit: <QUIT>.
error 45
The receive buffer of the Titrino contains an
incomplete string (missing L
F
). Transmission of the
Titrino is thus blocked.
Exit: Send
L
F
or <QUIT>.