Metrohm 799 GPT Titrino User Manual

Page 227

Advertising
background image

6.2 Pin assignment of the "Remote" socket

799 GPT Titrino

223

Outputs
































pin 17 (Output 3)

pin 3 (Output 4)

pin 16 (Output 5)

pin 1 (Output 6)

pin 2 (Output 7)

pin 6 (Output 8)

pin 7 (Output 9)

pin 8 (Output 10)

pin 13 (Output 11)

pin 19 (Output 12)

pin 20 (Output 13)

End of determination

EOD

L4 in TIP

Error,

active with errors

Activate pulse, see page 224.

L6 in TIP

Pulses for recorder (t

p

=150 µs)

10 000 per buret cylinder

Remote box active

not used

Monitoring, L10

Monitoring, L11

Monitoring, L12; sample size out of limits

Monitoring, L13; result out of limits

For all outputs:

V

CE0

= 40 V

I

C

= 20 mA

t

Pulse

> 100 ms

Functions see page 224

Voltage

pin 15

pin 14

pin 25

I ≤ 200 mA

0 V: active

5 V: inactive

Ordering numbers for plug:

K.210.9004 (shell) and K.210.002

No liability whatsoever will be accepted for damage caused by im-

proper interconnection of instruments.

0 V

+5 V