6 setting up the swing heads and robotic arms, 1 positioning the swing heads, Setting up the swing heads and robotic arms – Metrohm 864 Robotic Balance Sample Processor User Manual

Page 33: Positioning the swing heads

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

864 Robotic Balance Sample Processor

■■■■■■■■

25

3.6

Setting up the Swing Heads and robotic arms

3.6.1

Positioning the Swing Heads

In order to ensure that the robotic arms can be mounted as easily as pos-
sible, the Swing Head drives and/or the drive discs must be moved into a
defined position.

1 Initialize the Sample Processor

In tiamo

™ click the hand symbol for manual control (on the left-

hand sidebar).

Under the entry Sample changer/864_1, select the Tower 1.

In the right-hand window click Initialize rack.

All instrument components are moved to the starting position.

2 Prepare the Swing Head

Click the tab Move.

Under Robotic arm position click the button Arrow left until
the drive disc of the Swing Head on tower 1 does not move any
more.

Now the robotic arms can be mounted.