3.17 Adjusting the rack and the robotic arm

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

864 Robotic Balance Sample Processor

The last position on the sample rack is conceived as Adjusting posi-
tion
.

3 Define the special beaker on the 6.2068.010 sample rack

■

Select the lowest line (Special beaker 16) and click [Edit].

■

Under Rack position, select 61.

■

Close the dialog window with [OK].

■

Close the rack data table with [OK].

Moving to the positioning reticle

Proceed as follows:

1 Open the manual control

■

In the Manual control in tiamo

™, click the tab Move.

2 Move to special beaker 16

■

Under Rack position, select the target position Special beaker
16
and click [Start].

It is also possible to specify the target position as absolute rack posi-
tion. For the 6.2068.010 sample rack, the adjusting position is the
no. 61.

3 Move to the adjusting position

■

Unscrew the FEP tubing from the adapter on the robotic arm.

■

Insert the PEEK capillary supplied from above into the adapter.
The capillary must protrude below from the adapter by around
0.5 cm (see following illustration)

■

Under lift position enter 100 mm as target position and click
[Start].