Metrohm 847 Lab Link User Manual
Page 16

2 Installation
•
Advanced
Special adaptations to the local network surroundings can be carried out
here. These apply immediately for all Lab Link connections after confirmation
with [OK].
Caution!
You should only alter these settings when you have detailed knowledge of
the corresponding functions. Faulty timing settings may cause malfunction-
ing.
1. Under Log file level select the degree of detail for the information in the
Log file.
Standard
value: low, Selection: low, medium, high
Communication with the connected Titrando systems can be recorded in
a log file. This function may be particularly useful during startup for
searching for errors. The log file is deleted every 60 operating minutes or
after restart of the Lab Link Server; recording then starts again from the
beginning. The file is saved in the installation directory of the Lab Link
Server under bin\log.
12
847 USB Lab Link / Lab Link Server