Metrohm 847 Lab Link User Manual
Page 42

4 Troubleshooting
Time for Periodic
status test is set too
high
If the Lab Link Server list con-
tains a large number of non-
existent connections for which
the option Auto connect (via
multicast) has been deactivated,
the ICMP timeout period must
be included when calculating
the time (see formula for Peri-
odic status test, p. 13). We
recommend that unnecessary
connections are deleted from
the list.
Connection has been
interrupted with
Connection/ Discon-
nect
Turn off the Touch Control, wait
a few seconds, then switch it
back on.
Lab Link Server:
Status "not con-
nected"
Connection has only
been established
unidirectional
Restart the Lab Link Server.
Turn off the Touch Control, wait
a few seconds, then switch it
back on.
Touch Control Error
message:
"026-130 Incompati-
ble Software"
The software versions
of the Lab Link Server
and the Touch Con-
trols are not compati-
ble
On the Titrando installation CD
you will find versions that are
compatible.
38
847 USB Lab Link / Lab Link Server