Figure 6, Mounting the needles – Metrohm 874 USB Oven Sample Processor User Manual
Page 23

■■■■■■■■■■■■■■■■■■■■■■
3 Installation
874 Oven Sample Processor
■■■■■■■■
15
Mounting the injection and outlet needle
6.2816.080
6.2049.040/6.2049.050
6.2816.070
1
2
3
Figure 6
Mounting the needles
Mount the needles as follows
1 Screw the needle holder (6.2049.040 or 6.2049.050) into the distri-
butor on the guide head.
2 Screw the 6.2816.080 outlet needle onto the Luer connector of
the needle holder.
3 Carefully guide the 6.2816.070 injection needle into the opening
of the distributor from above and allow it to drop down.
Note
Take care to ensure that the white PTFE seal is positioned securely
on the needle.
4 Screw the 6.1805.470 FEP tubing by hand onto the opening of
the distributor. Screw the other end of the tube into the opening of
the gas outlet on the upper side of the tower.