Metrohm 850 Professional IC Anion MSM-HC MCS Prep 2 User Manual

Page 65

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

850 Professional IC –Anion – MSM-HC – MCS – Prep 2

■■■■■■■■

55

Note

The column input capillary is already threaded into the capillary recesses
of the column thermostat at the time the instrument is first delivered.
The following installation instructions need not be carried out at the
time of initial installation.

Threading capillaries

1 Introduce column input capillary into the column thermostat via a

suitable capillary feed-through (28-1).

2 Slide the column input capillary from below into the outer of the two

capillary recesses (28-2). Slide it through under the holder plate until
it emerges again at the top.

3 Carefully bend the column input capillary downward and slide from

from above to below through the inner capillary recess until it
emerges at the lower edge of the holder plate.

4

Note

The columns (guard column and separation column) may only be
installed after the initial start-up (see Chapter 4.1, page 75).

Before initial start-up:
Fasten the 6.2744.040 coupling with a 6.2744.010 pressure
screw to the end of the column input capillary.

After initial start-up:
Fasten the guard column (if used) or the separation column with a
6.2744.010 pressure screw to the end of the column input capil-
lary.

Advertising