3 Installation
940 Professional IC Vario ONE/HPG (2.940.1140)
■■■■■■■■
33
3.15
Installing the high-pressure gradient module
The outputs of the two high-pressure pumps are connected to the injec-
tion valve's eluent input using the mixing coil for high-pressure gradients
(6.2758.000).
Connect the two pieces of eluent aspiration tubing to an eluent bottle
(see Chapter 3.7, page 22).
3.16
Connecting the instrument to a computer
NOTE
The instrument must be switched off when being connected to a com-
puter.
Accessories
For this step you need the following accessories:
■
USB connecting cable (6.2151.020)