Metrohm 850 Professional IC Detector - iDetector User Manual
Page 17

■■■■■■■■■■■■■■■■■■■■■■
2 Installation
2.850.9010 IC Conductivity Detector
■■■■■■■■
9
NOTE
With the instruments of the 850 Professional line of instruments, up to
two detectors can be placed and connected.
Placing the detector in the instrument
1 Removing the knurled screws
■
Unscrew knurled screws on the rear panel.
■
Optional with 850 instruments: Remove the handle if it is still fas-
tened to the instrument.
2 Removing rear panel
■
Remove rear panel.
3 Positioning the detector
■
Put the detector through the opening and position it on the sup-
port surface intended for this purpose and slide it as far as the
limit stop to the front.
4 Replacing rear panel
■
Insert the detector cable in a cable feed-through on the rear
panel.
■
Place the detector outlet capillary in a suitable capillary feed-
through.
■
Replacing the rear panel.
The detector cable (3-1) and the detector outlet capillary (3-2)
should freely hang out of the cable or capillary feed-through.
5 Tightening the knurled screws
■
Optional with 850 instruments: Reattach the handle higher up
and use it as a holder for MPaks.
■
Tighten the knurled screws.