3 connecting the detector, Connecting the detector – Metrohm 850 Professional IC Detector - iDetector User Manual
Page 18

2.3 Connecting the detector
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
2.850.9010 IC Conductivity Detector
2.3
Connecting the detector
Connecting the detector cable and the detector outlet
capillary
CAUTION
The instrument must be switched off when connecting a detector.
1 Connecting the detector
■
Connect the detector connection cable (3-1) to the detector con-
nection socket Detector 1.
NOTE
Recommendation: In AnCat systems with 2 detectors: anions to
Detector 1, cations to Detector 2.
2 Connecting the detector outlet capillary
■
Guide the detector outlet capillary (3-2) into a sufficiently large
waste container and fasten it there.
NOTE
The detector outlet capillary must be free of blockages (the mea-
suring cell is tested to 5 MPa = 50 bar backpressure).
NOTE
The detector outlet capillary must not be shortened!
Shortening the detector outlet capillary leads to increased noise!