Page 22 – HF scientific Micro200BW Turbidimeter - 0-100 NTU User Manual
Page 34

MICRO 200 BW 0-10 & 0-100 NTU (05/05)
Rev. 3.4
3.2.6 Monitor Setup - Continued
By pressing in the Monitor Setup menu.
In this menu the baud rate can be changed to meet your requirements using the
keys. Buad rate of 300, 600, 1200, 2400, 4800 or 9600 may be selected.
By pressing in the Communications menu.
In this menu the serial printer mode can be turned On or Off. If the printer is selected On
and report is Off then the serial port will output the current reading in real time which is an
update to the output every second. Choose this mode if your sending the data directly to
a data logger or other data collection device. You can change the printer to On or Off by
using the
keys.
F4
F2
By pressing in Monitor Setup menu.
In this menu the 4-20mA output can be fine tuned to your requirements. You can change
either end or both ends of the 4-20mA output, using the keys.
Only the 20mA end of the scale is highlighted here. The menu highlighting the 4mA end
is intentionally not shown.
F3
By pressing again.
In this menu the 0-10 volt output can be fine tuned to your requirements. You can change
either end or both ends of the 4-20mA output, using the keys. Only the
volt end of the scale is highlighted here. The menu highlighting the 10 volt end is
intentionally not shown.
F3
The following menus will have no pupose
unless you have one of the optional serial ports
installed.
By pressing in the Monitor Setup menu.
In this menu the digital to analog converter can be set to output a 4-20mA signal or a 0-10
Volt signal using the
keys. The D to A automatically calibrates to the
upper and lower limits you selected in the Display Parameters menu.
F2
Page 22
MONITOR SETUP
10NTU TURBIDITY
TIME/DATE
1
∅:∅∅ MAR ∅4 1998
SELECT D TO A
4-2
∅MA
CURRENT/VOLTAGE ADJUST
4.
∅∅ 2∅.∅∅ ∅.∅∅ 1∅.∅∅
COMMUNICATIONS...
VER:M200NTUBW-8A23E
VER:M200NTUBW-8A23E
MONITOR SETUP
10NTU TURBIDITY
TIME/DATE
1
∅:∅∅ MAR ∅4 1998
SELECT D TO A
4-2
∅MA
CURRENT/VOLTAGE ADJUST
4.
∅∅ 2∅.∅∅ ∅.∅∅ 1∅.∅∅
COMMUNICATIONS...
MONITOR SETUP
10NTU TURBIDITY
TIME/DATE
1
∅:∅∅ MAR ∅4 1998
SELECT D TO A
4-2
∅MA
CURRENT/VOLTAGE ADJUST
4.
∅∅ 2∅.∅∅ ∅.∅∅ 1∅.∅∅
COMMUNICATIONS...
VER:M200NTUBW-8A23E
COMMUNICATIONS
10NTU TURBIDITY
BAUD
96
∅∅
PRINTER
ON
ADDRESS
1
REPORT
4HR
VER:M200NTUBW-8A23E
COMMUNICATIONS
10NTU TURBIDITY
BAUD
96
∅∅
PRINTER
ON
ADDRESS
1
REPORT
4HR
VER:M200NTUBW-8A23E