IRON

1,10-PHENANTHROLINE METHOD

CODE 3668

QUANTITY

CONTENTS

CODE

15 mL

*Acid Phenanthroline Indicator

*2776-E

5 g

*Iron Reducing Reagent

*2777-C

1

Spoon, 0.1 g, plastic

0699

*WARNING: Reagents marked with a * are considered hazardous substances. Material Safety Data Sheets (MSDS) are supplied for

these reagents. For your safety, read label and accompanying MSDS before using.

APPLICATION:

Drinking, surface and saline waters; domestic and industrial wastes.

RANGE:

0 – 6.0 ppm Iron

METHOD:

Ferric iron is reduced to ferrous iron and subsequently forms a colored complex with bipyridyl

for a quantitative measure of total iron.

SAMPLE HANDLE

& PRESERVATION:

The sample container should be cleaned with acid and rinsed with deionized water. Addition

of acid to adjust the sample to pH 2 - 3 will prevent deposition of iron on the container walls.

Samples should be analyzed as soon as possible.

INTERFERENCES:

Strong oxidizing agents, cyanide, nitrite, and phosphates; chromium, zinc in concentrations

exceeding 10 times that of iron; cobalt and copper in excess of 5 mg/L, and nickel in excess of

2 mg/L. Bismuth, cadmium, mercury, molybdate, and silver precipitate phenanthroline.

PROCEDURE

1.

Rinse a clean colorimeter tube (0967)with sample water. Fill to the 10 mL mark with sample.

2.

Select setting 2 on the “Select Wavelength” knob and press the “30 Second Read” button.

3.

Insert the tube into the colorimeter chamber and adjust to 100% T with the “Set Blank” knob. This is the 100% T

blank.

4.

Remove the tube from colorimeter. Use the 0.1 g spoon (0699) to add one measure of *Iron Reducing Reagent

(2777). Cap and invert the tube 15 – 20 times to mix.

5.

Remove the cap and add 6 drops of *Acid Phenanthroline Indicator (2776). Cap and invert the tube 3 – 4 times to

mix reagents. Wait five minutes for maximum color development.

6.

After five minutes, press the “30 Second Read” button and insert the tube into colorimeter chamber. Record %T as

soon as the reading stabilizes.

7.

Consult the calibration chart to determine iron concentration in parts per million (ppm).

DC1600 IRON ~ 1,10-PHENANTHROLINE CALIBRATION CHART

%T

9

8

7

6

5

4

3

2

1

0

90

0.02

0.03

0.05

0.07

0.08

0.10

0.11

0.13

0.15

0.17

80

0.18

0.20

0.22

0.24

0.26

0.28

0.29

0.31

0.33

0.35

70

0.37

0.39

0.41

0.43

0.46

0.48

0.50

0.52

0.54

0.57

60

0.59

0.61

0.64

0.66

0.69

0.71

0.74

0.76

0.79

0.81

50

0.84

0.87

0.90

0.93

0.96

0.99

1.02

1.05

1.08

1.11

40

1.15

1.18

1.21

1.25

1.29

1.32

1.36

1.40

1.44

1.48

30

1.52

1.57

1.61

1.66

1.70

1.75

1.80

1.85

1.91

1.96

20

2.02

2.08

2.14

2.21

2.27

2.34

2.42

2.49

2.57

2.66

10

2.75

2.84

2.94

3.05

3.16

3.28

3.41

3.56

3.71

3.88

0

4.07

4.29

4.53

4.82

5.16

5.58

6.13

47

Advertising