LaMotte DC1600 Colorimeter User Manual

Page 80

Advertising
background image

PROCEDURE

1.

Rinse a clean colorimeter tube (0967) with sample water. Fill to the 10 mL line with sample.

2.

Select setting 2 on “Select Wavelength” knob and press “30 Second Read” button.

3.

Insert tube into colorimeter chamber and adjust to 100%T with “Set Blank” knob. This is the %100T blank.

4.

Remove tube from colorimeter. Add 6 drops *Silica Reagent #1 (V-4466). Cap and invert to mix.

5.

Add 12 drops of *Silica Reagent #2 (V-4467). Cap and mix. Wait 5 minutes.

6.

At end of 5 minute waiting period, add 2 drops of *Silica Reagent #3 (V-4468). Cap and mix.

7.

Press “30 Second Read” button and insert tube into colorimeter chamber. Record %T as soon as reading stablilizes.

8.

Consult calibration chart to determine silica concentration in parts per million (ppm).

DC1600 SILICA ~ HIGH RANGE CALIBRATION CHART

%T

9

8

7

6

5

4

3

2

1

0

90

1

2

3

4

5

7

8

9

10

11

80

13

14

15

17

18

19

20

22

23

25

70

26

27

29

30

32

33

35

36

38

40

60

41

43

44

46

48

50

51

53

55

57

50

59

61

63

65

67

69

71

73

75

40
30
20
10
0

NOTE: To extend range to 150 ppm, perform a 2:1 dilution of water sample with silica free water. Perform test and

multiply result by 2.

80

Advertising