LaMotte DC1600 Colorimeter User Manual

Page 52

Advertising
background image

PROCEDURE

1.

Rinse a colorimeter tube (0967) with sample water. Fill to the 10 mL line with sample.

2.

Select setting 5 on the “Select Wavelength” knob and press the “30 Second Read” button.

3.

Insert tube into the colorimeter chamber and adjust to 100%T with the “Set Blank” knob. This is the 100%T blank.

4.

Remove tube from colorimeter. Use the 1.0 mL pipet (0354) to add 2.0 mL (two measures) of *Hardness Buffer

Reagent (4255). Swirl to mix.

5.

Add 2 drops of *Sodium Cyanide, 10% (6565). Cap and mix.

6.

Use the 0.5 mL pipet (0369) to add 0.5 mL of *Manganese Indicator Reagent (3956). Cap and mix.

7.

Press the “30 Second Read” button and insert tube into the colorimeter chamber. Record %T as soon as the reading

stabilizes.

8. Consult the calibration chart to determine the manganese concentration in parts per million (ppm).

NOTE: For best results, carry a reagent blank through the procedure. Treat 10 mL of demineralized water with the

test procedure. This is the Reagent Blank. Set the reading to 86.8%T, which corresponds to 0 ppm, then continue

with the unknown sample tests.

DC 1600 MANGANESE ~ LOW RANGE CALIBRATION CHART

%T

9

8

7

6

5

4

3

2

1

0

90

80

0.01

0.01

0.02

0.02

0.03

0.03

0.04

70

0.05

0.05

0.06

0.07

0.07

0.08

0.09

0.09

0.10

0.11

60

0.12

0.12

0.13

0.14

0.15

0.16

0.16

0.17

0.18

0.19

50

0.20

0.21

0.22

0.23

0.24

0.25

0.26

0.27

0.28

0.29

40

0.30

0.32

0.33

0.34

0.35

0.37

0.38

0.39

0.41

0.42

30

0.44

0.46

0.47

0.49

0.51

0.53

0.54

0.56

0.59

0.61

20

0.63

0.65

0.68

0.70

0.73

0.76

0.79

0.82

0.85

0.89

10

0

52

Advertising