LaMotte 2020i Turbidimeter User Manual

Page 21

Advertising
background image

10. Cap the tube. Wipe the tube

thoroughly again with a lint-free
cloth.

11. Open the meter lid. Insert the tube

into the chamber. Align the index
notch on the positioning ring with
the index arrow on the meter. Close
the lid.

12. Press

*IOK to select Scan Sample. Turbidity

( F )

* S c a n S a m p l e

q

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

13. Record the result.

T u r b i d i t y

( F )

0 . 5 4

N T U

* S c a n S a m p l e

q

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

14. Press

OFF to turn the meter off or

press

… to exit to a previous menu

or make another menu selection.

Note: The meter will remember the last scanned blank reading. It is not necessary to
scan a blank each time the test is performed. To use the previous blank reading,
instead of scanning a new one, scroll to Scan Sample and proceed. For the most
accurate results, the meter should be blanked before each test and the same tube with
tube positioning ring should be used for the blank and the reacted sample.