LaMotte 2020i Turbidimeter User Manual

Page 30

Advertising
background image

TROUBLESHOOTING GUIDE

Error Messages

Err1

Very Low battery. Replace battery or switch to AC power. Press back
arrow (

…) to back out. Scan sample again. Replace battery as soon

as possible.

Err2

The meter can not be calibrated outside of the allowable range of
the displayed reading. Confirm that standard was made correctly.
The displayed reading can only be adjusted to ±50% of the factory
calibration.

Err3

Meter can not be calibrated with a zero sample. Calibrate the meter
with a sample other than zero.

Err4

Processing error due to motion of suspended particles or submicron
air bubbles or opening/closing lid during readings. Scan sample
again until reading is obtained.

Err5

No blank reading. The meter has never been blanked for this test
factor. Blank meter.

Err6

Internal mathematical error. Re-blank the meter and rescan the
sample.

Err7

Configuration error. Call LaMotte Tech Service. Meter may have to
be returned for repairs.

low battery

Low battery. Change battery.

Troubleshooting

PROBLEM

REASON

SOLUTION

Lost in meter menus

Reset to factory default
settings. (see page 8)

Turn meter off. Hold down
*I

OK and press ON.

Release both buttons. Press
*I

OK to select the default

settings. Meter will turn off
and the factory settings will
be restored.

Usually large negative
or positive readings
when performing
calibration.

Incorrect standards used to
calibrate meter.

Use fresh 0.0 standard in
clean, labeled, vial. Reset
meter to factory default
settings. See procedure
above. Recalibrate meter.

Erroneous readings

Measurement was taken
with lid open.

Close lid. Read again.

>

on display

Over range.

The sample is outside of the
acceptable range. Dilute
sample and test again.

30

TROUBLESHOOTING

Advertising