LaMotte 2020i Turbidimeter User Manual

Page 27

Advertising
background image

12. Press

*IOK to select Scan Sample. Turbidity

( F )

* S c a n S a m p l e
1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

13. Observe the result.

T u r b i d i t y

( F )

0 . 5 4

N T U

* S c a n S a m p l e

q

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

14. Press

ˆ and then press *IOK to

select Calibrate.

T u r b i d i t y

( F )

0 . 5 4

N T U

* C a l i b r a t e

q

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

15. Use the

ˆ or ‡ to change the

highlighted digits on the display to
match the concentration of the
turbidity standard. Press

*IOK to

accept a digit and move to the
next digit.

C a l i b r a t e

0 0 . 5 4

q, Q, p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

C a l i b r a t e

0 0 . 5 4

q, Q, p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

C a l i b r a t e

0 0 . 5 4

q, Q, p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

C a l i b r a t e

0 0 . 5 0

q, Q, p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

ANALYSIS & CALIBRATION

27

0

4

5

0

0