LaMotte 2020i Turbidimeter User Manual

Page 24

Advertising
background image

Turbidity Standards

Only use AMCO or formazin standards with the 2020. StablCal

â standards below 50

NTU should not be used to calibrate the 2020. The diluent used in StablCal

â

standards has a different refractive index than traditional formazin standards and will
affect the results. The concentration of the calibration standard should be similar to the
expected concentration of samples that will be tested. The following standards are
available from LaMotte Company:

1480

0 NTU Standard, 60 mL (EPA and ISO)

1484

1 NTU Standard, 60 mL (EPA)

1481

1 NTU Standard, 60 mL (ISO)

1485

10 NTU Standard, 60 mL (EPA)

1482

10 NTU Standard, 60 mL (ISO)

1486

100 NTU Standard, 60 mL (EPA)

1483

100 NTU Standard, 60 mL (ISO)

CALIBRATION PROCEDURE

1. Press

*ION to turn the meter on.

1 . 3

2. Press

*IOK to select Measure.

M a i n M e n u
* M e a s u r e

D a t a L o g g i n g
O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

3. Rinse a clean tube (0290) three

times with the blank.

Below 1 NTU – The meter should be
blanked with a 0 NTU Primary
Standard or prepared turbidity-free
(<0.1 NTU) water. For the most
accurate results, use the same tube
for the blank and the sample.

24

ANALYSIS & CALIBRATION