Inchhg / kpa / bar selections – Amico Valve Alarm Combo Unit User Manual

Page 15

Advertising
background image

InchHg

KPA

BAR

Factory Default - InchHg

For InchHg mode, set the switch #4 to the ON position. The LED indicating InHg located next to

the vacuum source reading will illuminate.

For KPA mode, set the switch #4 to the OFF position and the switch #9 to the ON position. The

LED indicating KPA located next to the vacuum source reading will illuminate.

For BAR mode, the KPA indicating source must be changed to BAR by use of a label. Set the

switch #4 to the OFF and the switch #9 to the OFF position. The LED indicating BAR located

next to the vacuum source reading will illuminate.

Common Settings For Pressure and Vacuum

Repeat Alarm Enable/Disable

Factory Default - Disable

Disable

Set switch #5 to the OFF position to disable the repeat alarm.

NOTE: When the repeat alarm function is disabled, the alarm will not repeat.
Enable

Enable Mode: (Factory Default 30 min, when enabled).

Set switch #5 to the ON position.

PLEASE NOTE: The module with the lowest set Repeat Time is the one that controls the Repeat

Time. For example, if one module is set for 5 minutes and one for 30 minutes and both are Repeat

Alarm enabled, the alarm will now repeat every 5 minutes.

Setting Factory Default

To quickly reset the module (Pressure or Vacuum) to the factory default settings as follows:

• Pressure: High set-point 60 Psi, Low set-point 40 Psi.
• Nitrogen and HP Air: High set-point 195 Psi, Low set-point 140 Psi.
• Vacuum: Low set-point 12 inchHg.
• No Repeat alarm, but set for 30 min..

1. Set switch #8 to the ON position.

2. Turn the power off (wait 5 seconds) then back on.

3. Set switch #8 to the OFF position.

The module is now in the default mode.

InchHg

KPA

BAR

InchHg / KPA / BAR Selections

www.amico.com

15