System Power Supply

The System Power Supply has been pre-installed into the back box assembly. The System Power Supply

converts the AC voltage supply to the alarm into two voltages: 5 VDC (regulated) required by the

microprocessor hardware and 15 VDC (unregulated) required by the buzzer and the LED’s. This unit also

contains the main ON/OFF power switch, the transformer, the heat sink, the main fuse and fuse cover, the

rectifying circuitry, the terminal blocks and the low voltage DC power cable for connecting this unit to the

annunciator module. The System Power Supply can be easily removed and reinstalled by unscrewing it from

the back box.

Annunciator Module

The Annunciator Module contains the buzzer, a “Power On” LED as well as the “TEST” and

“ALARM MUTE” buttons. The function of the “TEST” button is to verify that the buzzer and

all the LED’s are in working condition. An alarm will be heard when this button is pushed

and all the LED’s will light up. When the button is released, the alarm will silence. The

“ALARM MUTE” button is used to silence an alarm that has occurred. This module also

contains a fail-safe relay that de-energizes when the buzzer is activated. This relay can be

used with the Amico remote buzzer, for applications requiring a remote audible alarm

(see Appendix B), for connection to another Amico alarm or a building management

system.

Alarm Valve Display Module

The module provides a digital display of the actual pressure/vacuum of a gas being

monitored. In addition, a gas trend indicator bar with High and Low alarms are displayed.

The trend bar has two coloured LED’s: GREEN for Normal and RED for High and Low

conditions.

Each display module contains a gas specific, colour coded label (USA or ISO colours are

available). The Display Module is field adjustable for pressure/vacuum settings, repeat

alarm and unit of measure. Dry contacts for high and low alarms are available for remote

monitoring of each module.

Sensor Module

The Sensor Module contains the transducer which converts the gas/vacuum pressure

source into a digital signal that is displayed on the display module. The sensor module

shall be housed in an anodized aluminum and nickel-plated brass enclosure to act as an

interference barrier. It is also temperature compensated.

Each sensor is clearly labeled and colour coded for the gas or vacuum being monitored.

The sensor module contains a gas specific DISS fitting to ensure correct connection of

the proper sensor to the respective gas. Each sensor has been factory calibrated by a

computer for the specific gas shown on the sensor housing. If it is not connected to the

appropriate gas display module, an error message (EO2) will be displayed.