Installation of alarm – Amico Valve Alarm Combo Unit User Manual

Page 9

Advertising
background image

For Local Sensor Only

If the sensors are to be mounted locally (inside the valve box), the pipe stubs must be connected to the

pipeline. Using silver-brazing techniques, connect each pipe stub to its appropriate gas or vacuum while

ensuring that the bottom of the pipe stub is wrapped with a damp cloth. BE CAREFUL not to damage the DISS

check-valve by overheating. When the brazing of pipe stubs has been completed, the system can be pressure

tested.

Standing Pressure Test

Perform a standing pressure test on the piping system as per NFPA-99 “Health Care Facilities” or

CSA-Z305.1 “Non-flammable Medical Gas Piping Systems”. Inspect all joints for leaks and make certain each gas

is piped to a correspondingly labeled gas service.

Frame/Module Assembly
Step #1

Remove the frame/valve window from its protective box

Step #2

Attach the Sensor wire to the display module (Sensor +, -)

Step #3

Close the valve window

Installation of Alarm