Technical Specifications
Supply Voltage:
115 to 220 VAC - 50 to 60 Hz
Current Draw:
1 Amp. Max.
Fuse (1/4 * 1-1/4):
Fast Blow 1 Amp.
Cable Requirements
AVC Display Module to Remote Sensor:
Distance:
Maximum 2,500 feet (750 m)
Cable:
Belden # 8451 or equivalent.
#22 gauge shielded, twisted pair (For multiple sensors a multi-conductor shielded
twisted pair cable should be used).
Signal:
30
VDC
- 1.0
Amps.
60
VDC
- 0.3
Amps.
125 VAC
-
0.5 Amps.
AVC Display Module to Master Module
Distance:
Maximum 10,000 feet (3,000 m)
Cable:
Minimum #22 gauge wire (does not have to be shielded, twisted pair)
Signal:
5 VDC, < 5 µA