Technical Specifications

Supply Voltage:

115 to 220 VAC - 50 to 60 Hz

Current Draw:

1 Amp. Max.

Fuse (1/4 * 1-1/4):

Fast Blow 1 Amp.

Cable Requirements

AVC Display Module to Remote Sensor:

Distance:

Maximum 2,500 feet (750 m)

Cable:

Belden # 8451 or equivalent.

#22 gauge shielded, twisted pair (For multiple sensors a multi-conductor shielded

twisted pair cable should be used).

Signal:

30

VDC

- 1.0

Amps.

60

VDC

- 0.3

Amps.

125 VAC

-

0.5 Amps.

AVC Display Module to Master Module

Distance:

Maximum 10,000 feet (3,000 m)

Cable:

Minimum #22 gauge wire (does not have to be shielded, twisted pair)

Signal:

5 VDC, < 5 µA