Bio-Rad Experion DNA Analysis Kits User Manual

Page 20

Advertising
background image

16

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

3. Examine the separation of at least one sample. Click on the sample name in the project tree or on

the lane in the virtual gel to view the electropherogram, which should have the following features (for
example, Figure 4.3):

n

Two resolved marker peaks

n

Sample peaks located between the upper and lower markers

n

Flat baseline

If the upper and lower marker peaks are not properly assigned, you may need to include or exclude

peaks, or manually set the marker (see Sections 6.2 and 6.3).

4. Evaluate the data analysis performed by Experion software (see Chapter 4, Data Analysis). If

necessary, change the analysis settings and parameters — including quantitation methods — by
following the instructions in Chapters 5 and 6.

Fig . 3 .6 . Separation of the DNA ladder . Top, DNA 1K ladder. Bottom, DNA 12K ladder. Note the flat baseline and well-resolved peaks. All
identified peaks are numbered, and the lower and upper markers are indicated by green asterisks (*).

DNA ladder peaks

Baseline

Upper
marker

Lower
marker

DNA ladder peaks

Baseline

Upper
marker

Lower
marker