Fig . 6 .4 . Manual integration mode indicates the beginning and end of peak baselines with blue dots .

6.8 Manual Peak Integration

This function enables definition or refinement of peak areas for the ladder and all sample lanes.

1. Select a sample and click or select Analysis > Manual Integration to start manual integration.

2. Adjust the baseline by clicking and dragging the dots at the ends of the peaks (Figure 6.4).

If a peak in the electropherogram is not detected by the software, right-click on that peak and select
Add Peak. Zoom in on the peak for a better view of the baseline.

3. Save any changes by selecting File > Save Run As (use a new file name), or revert to the original

baseline and integration by clicking .

Manual integration applies to individual sample wells. Moving back to automated integration loses
changes made during manual integration unless those changes are saved.