Main component specifications – Bio-Rad Personal Molecular Imager™ (PMI™) System User Manual
Page 29
![background image](/manuals/600300/29/background.png)
25
Appendix 1
Personal Molecular Imager Specifications
System Technical Specifications
Specification
Linear dynamic range
1:65535
Pixel resolution
50, 100, 200 or 800 µm, selectable
Image resolution
2 line pairs/mm or 250 µm
Pixel density
16-bit (0–65,535)
Signal decay
32
P 50% retention in 24 hr
Scanning area
35 x 43 cm
Main Component Specifications
Laser Scanner
Specification
Dimensions
59 x 69 x 30.25 cm
Construction
Plastic molded housing, Aluminum mounting
plates
Excitation source
532 nm diode pumped solid state laser 25 mW
Weight
30 kg
Electrical
Maximum power
650 W
Input voltage range
90–260 VAC, 50–60 Hz
Fuses
No user-serviceable fuses
Operating Requirements
10–32°C, 30–80% humidity
Storage Requirements
0–60°C, 10–90% humidity
Screen Eraser-K
Specification
Dimensions
48 x 35 x 57 cm
Construction
Molded plastic
Weight
8.6 kg
Electrical
Input voltage range
100/120 VAC, 50–60 Hz
220/240 VAC, 50–60 Hz
Fuses
6.3 Amp (100/120 VAC)
3.15 Amp (220/240 VAC)
Illumination
4 x 15 Watt user-replaceable fluorescent bulbs
Operating Requirements
10–32°C, 30–80% humidity
Storage Requirements
0–60°C, 10–90% humidity