8

1. Plan Plate Layout

Determine the total number of wells in the experiment using the Plate
Layout Template on page 34 or the Plate Formatting tab in
Bio-Plex Manager

. A suggested plate layout is shown in Figure 2,

with all conditions in duplicate.

1. Assign standards to columns 1 and 2, with the highest concentration

in row A and the lowest concentration in row H.

2. Assign the blank to wells A3 and A4. The blank should consist of your

chosen standard diluent HB. Note that Bio-Plex Manager automatically
subtracts the blank (B) MFI value from all other assay wells.

3. User-specified controls, as well as the controls supplied in premixed

kits, are assigned to wells in columns 3 and 4.

4. The remainder of the plate is available for samples.

5. Once the total number of wells is known, you can calculate the

required volumes of beads, detection antibody, and streptavidin-PE.
Use Tables 6–7, 9–10, and 11, respectively, or the Calculation
Worksheet on pages 35–36.

Legend

S Standard

B Blank

X Samples

C Controls

Fig. 2. Suggested plate layout. For detailed instructions on plate
formatting in Bio-Plex Manager, see the Read Plate section.