Previous versions of bio-plex manager software, Luminex xponent software – Bio-Rad Human MMP and TIMP Assays User Manual

Page 30

Advertising
background image

28

Removing Outliers
Outliers are identified as standard data points that do not meet accuracy
or precision requirements and should be considered invalid when
performing curve fitting. As such, they should be removed to generate a
more realistic and accurate standard curve. This may result in an extended
assay working range and allow quantitation of samples that might
otherwise be considered out of range.

In Bio-Plex Manager software version 6.0 and higher, outliers can be
automatically removed by selecting the Optimize button in the Standard
Curve window. In Bio-Plex Manager software 5.0 and earlier versions,
outliers also can be manually selected in the Report Table. Visit online Help
to learn more about the standard curve optimizer feature and how outliers
are determined.

Previous Versions of Bio-Plex Manager Software

For instructions on using previous versions of Bio-Plex Manager software,
please contact Bio-Rad Technical Support.

Luminex xPONENT Software

Although guidelines are provided here, consult the xPONENT software
manual for more details. Perform a system initialization with Luminex’s
calibration and performance verification kit, as directed by Luminex. Select
Batches to set up the protocol and follow the information under Settings.

Note: The instrument settings described below apply to Luminex
100/200 and FLEXMAP 3D or Bio-Plex

®

3D instruments. For the

Bio-Plex

®

MAGPIX

reader, use the default instrument settings.

1. Select MagPlex as the bead type for magnetic beads, which

automatically sets the DD gates.

2. Volume = 50 µl.

3. Refer to Table 12 to select the appropriate PMT setting for

your instrument.

4. Plate name: 96-well plate.