Prepare wash method, Calibrate system, Setting up the bio-plex pro wash station – Bio-Rad Human MMP and TIMP Assays User Manual

Page 12: Setting up the bio-plex handheld magnetic washer

Advertising
background image

10

Calibrate System

1. Select Calibrate

and confirm that the default values for CAL1

and CAL2 are the same as the values printed on the bottle of Bio-Plex
calibration beads. Use the Bio-Plex system low RP1 target value.

2. Select OK and follow the software prompts for step-by-step
instructions for CAL1 and CAL2 calibration.

Note:

In Bio-Plex Manager version 6.1 and higher, startup, warm up,

and calibration can be performed together by selecting the Start up and
calibrate icon.

3. Prepare Wash Method

Bio-Plex Pro

assays are compatible with both magnetic separation and

vacuum filtration methods. However, for best results, we recommend
performing the assays in a flat bottom plate with magnetic separation.

Setting Up the Bio-Plex Pro Wash Station

The wash station should be primed before use. For more information, refer
to the Bio-Plex Pro Wash Stations Quick Guide (bulletin #5826).

1. Install the appropriate plate carrier on the wash station.

2. Use the Prime procedure to prime channel 1 with 1x wash buffer.

Setting Up the Bio-Plex Handheld Magnetic Washer

Place an empty flat bottom plate on the magnetic washer by sliding
it under the retaining clips. Push the clips inward to secure the plate.
Make sure the plate is held securely. If needed, the clips can be adjusted
for height and tension. For detailed instructions, refer to the user guide
(bulletin #10023087).

Table 3. Summary of compatible wash stations and plate types.
Wash Method

Wash Station

Assay Plate

Magnetic separation

Bio-Plex Pro

Flat bottom plate

Bio-Plex

®

handheld magnetic washer