23

Bio-Plex Manager software versions 6.0 and higher contain protocols for
most Bio-Plex

®

assays. Choose from available protocols or create a new

protocol. To create a new protocol, select File, then New from the main
menu. Locate and follow the steps under Protocol Settings.

1. Click Describe Protocol and enter information about the assay

(optional).

2. Click Select Analytes and create a new panel. Visually confirm the

selected analytes and proceed to step 3.

a.

Click

Add Panel

in the Select Analytes toolbar. Enter a new

panel name. Select Bio-Plex Pro Assay Magnetic from the
assay dropdown list. If using Bio-Plex Manager version 5.0 or lower,
select MagPlex from the assay dropdown list.

b.

Click

Add. Enter the bead region number and name for the first

analyte. Click

Add Continue to repeat for each analyte in the

assay. Refer to the bead regions in parentheses ( ) listed on the
peel-off label provided with the standards.

For reference, bead regions are shown in Table 13.

c.

Click

Add when the last analyte has been added and click OK to

save the new panel.

d. Highlight analytes from the Available list (left) and move to the

Selected list (right) using the Add button. To move all analytes at
once, simply click Add All.

e. If some of the analytes need to be removed from the Selected

list, highlight them and select Remove. If desired, it is possible to
rename the panel by clicking Rename Panel and entering a new
panel name.